Stay up-to-date with RSS.CO2 Incubator Technical Bulletins

ID Title Expand All
ATB0091
203.06KB
DH-CO2 Automatic Tank Switch Installation Instructions

The Automatic CO2 Tank Switch (internal) has been designed to provide reliable service to perform the critical backup CO2 tank switch for the DH Autoflow incubator. The Tank Switch continuously monitors the selected tank. This bulletin describes how to install the automatic tank switch in NU-5500/E/G model incubators.

ATB0198
144.78KB
INSTALLATION INSTRUCTIONS FOR DIVIDED GLASS DOORS
ATB0085
18.84KB
NU-1590 CO2 Single Stage Regulator Installation Instructions
ATB0045
19.22KB
Optional Feature (I11) Internal Coil for Chilled Water

The NuAire AUTOFLOW Series of water-jacketed CO2 incubators are available with an internal water-jacket cooling coil for ambient temperature or slightly below ambient temperature cell culture. This option for NuAire Incubators would have temperature capability to 18.0°C. The internal water-jacket cooling coil, when connected to a recirculating chilled water bath (customer supplied), will provide accurate temperature control to +0.1°C with uniformity being +0.2°C.

GTB0042
8.86KB
EMC Directive (89/336/EEC)
GTB0098
18.91KB
NU-4950/E/D/G N2 Gas Consumption
GTB0007
135.27KB
NuAire Incubator Sterility Test

Incubator contamination is an inherent risk for all incubator manufacturers. Typical laboratories have thousands of airborne contaminants that may enter a tissue cell culture incubator during a door opening, and enter the perfect growth environment to survive and reproduce. This bulletin describes how to calculate inner chamber air sterility in a NuAire co2 incubator.