LabGard® ES NU-677 Class II, Type A2 Animal Handling Biosafety Cabinet

Personnel, Product, and Environmental protection provided by a NSF/ANSI listed Class II Biological Safety Cabinet. A large 12" (305 mm) access opening accommodates the largest mouse and rat cages.


NSF/ANSI 49   UL Canada United States Listing


Part #: NU-677

Features

The LabGard ES (Energy Saver) Model NU-677 Animal Handling Biological Safety Cabinet, which comes in three sizes, [4 ft., 5 ft. and 6 ft. (1.2 m, 1.5, and 1.8 m)] is a bench top animal handling biosafety cabinet that is designed to give you an optimum amount of control over product quality while simultaneously reducing the potential of exposing either lab personnel or the product to airborne biological or particulate chemical agents. The cabinet is designed for use in low to moderate risk-hazard research and drug preparation or product operations, and its Energy Saver DC ECM motor is used to maximize both energy efficiency and filter loading capacity.


Ergonomics

Convenient Access Opening

The cabinet is designed with a 12" (305 mm) access opening, which gives you more moveability in the cabinet and makes working within it convenient.


Flexible Usage

The animal handling biosafety cabinet was designed with flexibility in mind. It is height adjustable, which means you can avoid injuries and ensure your body is comfortable while you work in the lab. The cabinet is also built with casters so that it can be easily moved around the workspace. A built-in pass-through gives you the opportunity to do a safe room-to-room transfer, and a waste station offers access to a waste bin while you work inside of the cabinet. The flexible features of the LabGard cabinet allow you to maximize your workflow in the lab.


Safety

Laminar Airflow

Laminar Airflow is key within an animal handling biosafety cabinet for preventing cross contamination. Constant uniform air passes over the work surface in the LabGard ES Energy Saver Class II, Type A2 at an airflow velocity of 60 fpm (0.30 m/s), which helps keep products and lab animals safe.


Air Barrier

To protect anyone working in the lab, the animal handling biosafety cabinet has an aerodynamic airfoil on its front. This air barrier with an airflow velocity of 105 fpm (0.53 m/s) helps isolate the work area from the lab.


NSF Listed for Your Protection

NuAire wants to ensure your protection in the lab. The 4, 5 and 6-foot models of the LabGard Model NU-677 cabinet are NSF/ANSI 49 listed for biological containment protection, which means both you and your product will stay safe when you use them to work in the lab. NuAire regularly validates the listing, to ensure it is accurate.


Technology

Energy Saving Motor to Lower Operating Costs

The animal biosafety cabinet is outfitted with an ultra-high efficiency Energy Saver motor. This ES motor is intended both to extend filter life and to use less energy overall - helping you reduce operating costs. As an added benefit, the motor produces less heat, which means you will not need to use your HVAC system in the lab.


Specs

Catalog Number

Catalog Number

NU-677-400

Nominal 4 foot (1.2m)

NU-677-500

Nominal 5 foot (1.5m)

NU-677-600

Nominal 6 foot (1.8m)

Performance Specifications

  1. Personal Protection

  2. Product Protection


NSF/ANSI 49


NSF/ANSI 49


NSF/ANSI 49

NSF Std. No. 49 Class

Class II, Type A2

Class II, Type A2

Class II, Type A2

Style of Cabinet

Bench top/console w/basestand

Bench top/console w/basestand

Bench top/console w/basestand

Cabinet Construction

All welded stainless steel 16GA, Type 304 pressure tight design

All welded stainless steel 16GA, Type 304 pressure tight design

All welded stainless steel 16GA, Type 304 pressure tight design

Diffuser for Air Supply (Metal)

Non-flammable

Non-flammable

Non-flammable

HEPA Filter Seal Type:

Supply Filter-99.99% Eff. on 0.3 microns Exhaust Filter-99.99% Eff. on 0.3 microns


HEPEX Seal

Neoprene, Spring-loaded


HEPEX Seal

Neoprene, Spring-loaded


HEPEX Seal

Neoprene, Spring-loaded

Fumigation per NIH/NSF Procedure

Yes

Yes

Yes

Standard Services: Duplex Outlet

Two, Backwall

Two, Backwall

Two, Backwall

Optional Services: Gas Cocks 3/8" NPT Ultraviolet Light


Up to 3 ea. Sidewall One, Backwall


Up to 3 ea. Sidewall One, Backwall


Up to 3 ea. Sidewall One, Backwall

Cabinet Size Inches (mm): Height Maximum

Height Minimum Width

Depth (with Control Center and Armrest)


97-1/2 (2477)

79 (2007)

53-5/8 (1362)

32-3/4 (832)


97-1/2 (2477)

79 (2007)

65 5/8 (1667)

32-3/4 (832)


97-1/2 (2477)

79 (2007)

77 5/8 (1972)

32-3/4 (832)

Work Access Opening Inches (mm): Standard Opening Height/Optional Standard Inflow Velocity

Work surface Height Min/Max


12 (305)

105 FPM (.53 m/s)

24-1/2 (662) / 43 (1092)


12 (305)

105 FPM (.53 m/s)

24-1/2 (662) / 43 (1092)


12 (305)

105 FPM (.53 m/s)

24-1/2 (662) / 43 (1092)

Work Zone Inches (mm): Height

Width

Depth Measured at 12 inches (305) Window Height


28 (711)

46 3/8 (1178)

25 (635)


28 (711)

58 3/8 (1483)

25 (635)


28 (711)

70 3/8 (1788)

25 (635)

Viewing Window Inches (mm): Standard is tempered sliding glass

Fully closed to 20 (508) open

Fully closed to 20 (508) open

Fully closed to 20 (508) open

Required Exhaust CFM/CMH Standard/ Optional:

Variable Flow Canopy (NU-911) Fixed Flow Canopy (NU-907)


486-656 (626-1115)

526 (894)


611-761 (1038-1293)

651 (1106)


758-868 (1288-1475)

796 (1352)

Plant Duct Static Pressure Eng./Metric

0.05-0.1"/1.27-2.54mm

0.05-0.1"/1.27-2.54mm

0.05-0.1"/1.27-2.54mm

Heat Rejected, BTU, Per Hour (non-vented)

(vented)

2551

157

2669

178

2787

198

Electrical: Volts, AC 60 Hz

+Amps: Blower/Lights/Autolift Amps: Duplex

Amps: Total

12 ft. Power Cord (one)

U.L./U.L.-C Listed 115

6.5

3

15

12 GA - 3 Wire, 20A

U.L./U.L.-C Listed 115

6.8

3

16

12 GA-3 Wire, 20A

U.L./U.L.-C Listed 115

7.1

3

16

12 GA-3 Wire, 20A

* Crated Shipping Weight: Net Weight

800 lbs. /362 kg. 750 lbs. /340 kg.

865 lbs. /392 kg. 815 lbs. /369 kg.

930 lbs. /421 kg. 880 lbs. /398 kg.


* Crated shipping weight does not include weight for accessories or options

+ Based on cabinet with new filters running at 115VAC.

Download Resources
Reviews

Be the first to review this product

Write Your Own Review

Only registered users can write reviews. Please, log in or register