Products

Find a Field Service Technician

Hitachi Koki himac CS150NX Tabletop Micro Ultracentrifuge

Model CS150NX features a re-engineering drive shaft to produce speeds of 150,000 rpm and a RCF of 1,050,000 xg. Simple operations allow users to easily separate and purify proteins, DNA, RNA, viruses, as well as carbon nanotubes and nano-sized particles.


Be the first to review this product




Download Specifications

CS150NX Angled View

More Views

  • CS150NX Front View
  • CS150NX Front View
  • CS150NX Side View
  • CS150NX Top View
  • CS150NX Touch Screen Run Screen Shot
  • CS150NX Touch Screen Menu Screen Shot

Features
Specs

Technologies

Reliable Drive Unit

The CS150NX only takes 90 seconds to reach the world’s fastest speed 150,000rpm by himac original vacuumed drive unit. The Drive Unit is backed by a 5 Year Warranty.

Imbalance Protection System

Samples need to be balanced within 5mm by visual check only.  A non-contact imbalance sensor monitors vibration of the rotor and drive shaft. In the event of unexpected vibrations the Imbalance Protection System activates and ceases operation immediately.

Self-Locking Rotor System

Hitachi Koki himac CS150NX Table Top Micro Ultracentrifuge Self-Locking Rotor System Installing and removing a rotor has never been simpler. Place the rotor on the drive shaft and thats it. The rotor is automatically locked by centrifugal force. There is no need for screws, tools, or buttons.

Simplified Operation with Touch Screen Controls

Hitachi Koki himac CS150NX Table Top Micro Ultracentrifuge Touch Screen Controls A color touch screen LCD with graphic user interface simplifies user operation by using universal icons. High contrast text and buttons on a black color background makes user system recognition easy.

Quiet Operation of 45 dbA

Hitachi Koki himac CS150NX Table Top Micro Ultracentrifuge Quiet Operation Just louder than a whisper, the CS150NX can attribute its quiet operation to rigid-control drive system and sound absorbing structure design.

Easy Timer Setting and Actual Run Timer

The himac original RTC (real-time control) function makes timer setting easy.  Simply set the start time or finish time along with running time.  It is easier than conventional delay time setting timer.


The himac original Actual Run Timer starts when the setpoint speed is attained and excludes acceleration time from the set time.  It helps to precisely control the net run-time and obtain high-reproducibility separation.  Of course, conventional run timer is selectable.

User Administration with Password Protection

Control users by regiserting them to the system with different access levels and passwords.

 

 

Options

Biosafety Option

Hitachi Koki himac CS150NX Table Top Micro Ultracentrifuge Biosafety Option In order to prevent from exhausting bio-hazardous samples into the laboratory, a micro-filter can be assembled in a vaccum line as an option.

Model

CS150NX

Maximum Speed (rpm) 150,000
Maximum RCF (xg) 1,050,000 (S140AT rotor)
Speed Control Accruacy (rpm) ±50 (5,000 to max speed)
Maximum Capacity (nominal) 30ml x 6 tubes (S50A rotor)
Maximum Capacity (nominal) (Swinging Bucket Rotor) 7ml x 4 tubes (S50ST rotor)
Timer 1 min. to 99 hrs. and 59 min. with HOLD and RTC (real-time control) function
Vaccum System Oil rotary vaccum pump and oil diffusion pump
Rotor Temperature Control Range (°C) 0 to +40 (1 degree increment)
Rotor Cooling Method Thermo-module Cooling System (HFC free)
Screen Display Color LCD (touch-sensitive)
Rotor Setting Method Self-locking Rotor System (no tools required)
Operational Noise (dbA*) 45 *Measured 1 meter from front of unit
Maximum Heat Dissipation into Room (kW) 0.7 or Less
Exertior Dimensions (W x D x H) 23 ¼ x 22 ¾ x 16 ¼ in. (590 x 582 x 408 mm)
Floor Area 5½ ft² (33 ½ x 23¼ in.) [0.50 m² (850 x 590 mm)
Weight 214 lbs. (97 kg)

Power Requirements

AC 110 or 120V ±10%, 15A (50/60 Hz), Single Phase

AC 208, 220, 230, 240 V ±10%, 8A (50/60 Hz), Single Phase

   
Drive Unit Warranty 5 Years