| ID | Title Expand All |
|---|---|
|
ATB0147 52.14KB |
|
|
ATB0081 6.25MB |
NuAire airflow products may have optional storage trays installed. This bulletin describes the placement of these storage bins on several NuAire products. |
|
ATB0146 468.21KB |
The NU-NR797 Negative Pressure Recirculating Isolator exhausts roughly 5% of its total volume to maintain the Isolators negative pressure. It is strongly recommended, per OSHA, NIOSH and ASHP, that this Isolator exhaust volume be exhausted to the outside. NuAire recommends the use of a canopy transition. Canopy transitions, sometimes referred to as a thimble or air gap exhaust transition, provide an air gap between the exhaust efflux of the Isolator and the transition. The air gap essentially insulates the Isolator from potential variations of airflow in the plant exhaust system, particularly if multiple items are connected to the same exhaust blower system. This bulletin details how to install such transitions on NU-NR797 Pharmacy Isolators. |
|
GTB0042 8.86KB |
|
|
GTB0151 770.53KB |
This bulletin describes air quality testing requirements per USP797 with respect to the NuAire NU-NR797 pharmaceutical compounding aseptic containment isolator. |
|
GTB0152 1.12MB |
This bulletin describes air quality testing requirements per USP797 with respect to the NuAire NU-NTE797 pharmaceutical compounding aseptic containment isolator. |
|
GTB0150 707.69KB |
This bulletin describes air quality testing requirements per USP797 with respect to the NuAire NU-PR797 pharmaceutical compounding aseptic isolator. |