The following NuAire Units are currently listed to the NSF/ANSI Standard 49 -2002 and are required to be field tested to that standard:
Because of customer demand, NuAire has started manufacturing a 5-foot version of our NU-425 Biological Safety Cabinet. For years the NU-425 has only been available in a 2ft., 3ft., 4ft., or 6ft. widths, now the 5ft. is available. In addition, the models NU-437 and NU-440 can also be produced in a 5ft. wide size. These 5ft. units contain all the features, construction and operating characteristics of the parent NU-425's. All options available on the NU-425 product line are available on the NU-437 and NU-440 models.
All the models: NU-425-500, NU-437-500 and NU-440-500 are NSF / ANSI Standard 40 - 200 listed. The primary inflow measurement method is via the DIM (Direct Inflow Measurement device). The alternate means of measuring inflow is via the 3 inch constricted method (Reference NuAire Technical Bulletin TB0115). Down flow specification is 65 - 75 fpm. Inflow specification is 100 - 110 fpm. 10-inch window units exhaust 426 CFM, and the 8-inch window height units exhaust 340 CFM.
NuAire has produced a new Data Plate for its Biological Safety Cabinets that conforms to the
NSF/ANSI Standard 49 - 2002 format. Reference NuAire Technical Bulletin
Contact NuAire's Technical Service Department for additional information. You can also click on the NSF web site link for all NuAire listed models and specific unit information.