LabGard® ES AIR NU-543E Class II Microbiological Safety Cabinet

The LabGard® ES (Energy Saver) AIR NU-543E is a bench top model with base stand options. Model NU-543E holds TUV and NSF listings meeting industry standards EN 12469 and NSF/ASNI 49 at a 254 mm (10-inch) access opening. LabGard offers optimum product quality while reducing the potential for exposure of both product and personnel to airborne biological or particulate chemical agents in low to moderate risk-hazard research and drug preparation or product operations.


NSF/ANSI 49  EN12469   CE

Part #: NU-543E

Features

The LabGard® ES model NU-543E Class II Biosafety Cabinet is offered as a 230-volt cabinet with a 254 mm (10-inch) access opening, TUV and NSF listings in four unique sizes. Model NU-543E offers the very best in personnel, product, and environmental protection.


Safety and Performance

Industry Listings

Model NU-543E holds TUV and NSF listings meeting standards EN 12469 and NSF/ASNI 49 for biosafety cabinetry at a 254 mm (10-inch) access opening in nominal 0.9 meters (3-foot), 1.2 meters (4-foot), 1.5 meters (5-foot), and 1.8 meters (6-foot) widths.


Dynamic Air Barrier

An air barrier at the front opening of the cabinet provides personnel protection by creating an inflow velocity of 0.53 m/s (105 fpm).


True Laminar Airflow

Uniform air is provided from the supply filter at a download rate of 0.30 m/s (60 fpm) minimizing cross contamination.


Construction

A Size that Fits Your Workflow

Model NU-543E is offered in four unique sizes to meet your application and facility space. NU-543E is offered in nominal 0.9 meter (3-foot), 1.2 meter (4-foot), 1.5 meter (5-foot), and 1.8 meter (6-foot) widths.


Monolithic Design

The process starts with a single sheet of 16/18 gauge, Type 304 stainless steel that is bent to create the exterior walls of the cabinet. Two additional sheets are seam welded into place creating the ceiling and floor of the cabinet. A smaller sheet of stainless steel is then bent and welded into place to create the interior walls of the cabinet. The result is a solid single piece shell that eliminates the user of rivets, gaskets, and sealants.

NU-543E Construction Matters
 

Coved Interior

A sheet of 16/18 gauge, Type 304 stainless steel is bent to create the interior walls of the biosafety cabinet. The corners of the work zone are coved creating an easier surface to clean and maintain.


Filter Integrity Check

Model NU-543E features large H14 99.995% @ 0.025 MPPS HEPA filters that are tested using EN 1822-1 testing standards as prescribed by biosafety cabinetry standard EN12469. Exhaust and Supply HEPA filters feature more filter media equating to longer filter life and lower operating costs. HEPA filters are tested twice during the manufacturing process. Once before the filter is installed and again fitted in the biosafety cabinet before it ships. This integrity check process ensures you receive the safest biosafety cabinet out of the box with zero headaches. We believe in our process so much we included our filters in our warranty.


Technology

Ultra High-Efficiency Motor

Not only save on your energy bills but your total cost of ownership. A single ultra-high efficiency motor features more torque than previously used technologies extending the life of your filters.


FlowGard™ Electronic Control System (ECS)

Using the same technology as your service professional uses to certify your biosafety cabinet, FlowGard™ ECS features a dual thermistor airflow sensor (Intelliflow) to monitor the cabinet’s airflow performance. The ECS also automatically compensation for filter loading and line voltage variances maintaining airflow performance resulting in consistent safety conditions.


HEPEX™ Zero Leak Airflow System

A flexible plenum produces true laminar airflow over the work zone by evenly displacing air over the supply filter. The positive plenum is maintained under negative pressure, so any potential leak will be re-directed back into the plenum and pass through the supply or exhaust filter.


High-Efficiency Particulate Air Filters

Model NU-543E features H14 HEPA filters in the supply and exhaust airflow track. H14 filters are 99.995% efficiency at most penetrating particle size.


Ergonomics

A Larger User-friendly Work Zone

A large user-friendly work zone reduces clutter allowing you to create a safer work environment allowing for proper aseptic technique minimizing potential cross contamination.


Shorter Reach into the Work Zone

The shortest reach into the safe work area in the industry, model NU-543E features a 152 mm (6-inch) reach until you reach the work zone. The short reach help reduces stress on back, neck, and arms minimizing workplace injuries.


Warranty

Back by the Industry’s Best

6-Years Parts including Filters (International)

5-Years Parts including Filters and Labor (United States and Canada)


Options

Access Openings

Part Number Description
M444 203 mm (8-inch) Access Opening Safety Plate Sliding Glass
M442 254 mm (10-inch) Access Opening Safety Plate Sliding Glass
M446 203 mm (8-inch) Access Opening Safety Plate Sliding Glass, Glass Sidewalls
M445 254 mm (10-inch) Access Opening Safety Plate Sliding Glass, Glass Sidewalls

Supply Filter

Part Number Description
M440 H14 Metal Frame HEPEX Separatorless 99.995% @ 0.025 MPPS with Permanent Plenum

Exhaust Filter

Part Number Description
M344 H14 Metal Frame HEPEX Separatorless 99.995% @ 0.025 MPPS
M315 Redundant (Qty 2) H14 Metal Frame HEPEX Separatorless 99.995% @ 0.025 MPPS

Lighting


Part Number Description
E121 Fluorescent Lighting (230V)
E122 LED Lighting

Armrest

Part Number Description
M324 Heavy Duty Ergonomic PVC Armrest
M211 Stainless Steel Armrest

Built-in Options

Part Number Description
E21 Ultraviolet Light (230V)
E98 Blower Contact Wires
M262 IV Bar w/ 6 Hooks
M427 NSF Cord Pass-Thru Right Side Wall
M428 NSF Cord Pass-Thru Left Side Wall
M411 Stainless Steel Window Edge Protector
M387 Prop-up Work Tray
M316 Australian Test Method

Service Valves

Part Number Description
M198 Right Bottom Service Valve
M193 Left Top Service Valve
M194 Left Middle Service Valve
M195 Left Bottom Service Valve
M196 Right Top Service Valve
M197 Right Middle Service Valve

Service Couplings

Part Number Description
M205 Left Top Service Coupling
M206 Left Middle Service Coupling
M207 Left Bottom Service Coupling
M208 Right Top Service Coupling
M209 Right Middle Service Coupling
M285 Right Bottom Service Coupling

Plumbing

Part Number Description
M120 3/8" Soft Copper Pipe
M121 3/8" Black Pipe
M122 1/2" Rigid Copper Pipe

Side Panels

Part Number Description
M284 Side Panel without Plumbing (Qty 2)
M237 Side Panel w/ Plumbing Out Bottom (Qty 2)
M238 Side Panel w/ Plumbing Out Top (Qty 2)
M239 Side Panel w/ Plumbing Out Back (Qty 2)

Outlets

Part Number Description
E42 Australian - Type I Outlet (230V)
E41 UK - Type G Outlet (Current) (Qty 2)
E39 France - Type E Outlet (230V)
E43 Israel - Type H Outlet (230V)
E38 UK - Type D Outlet (Past) (230V)
E40 Schuko - Type F Outlet (230V) (Qty 2)
E44 Switzerland - Type J Outlet (230V)
E37 US - Type B Outlet (230V) (Qty 2)
E111 South Africa - Type M Outlet (230V)

Power Input

Part Number Description
E07 Junction Box
E62 Power Cord 230V

Seismic

Part Number Description
M181 Seismic Studs
M264 Seismic Brackets
M306 Seismic Brackets for Adjustable Base Stand

Specs


Catalog Number

Catalog Number

NU-543-300E

Nominal 3 foot (0.9m)

NU-543-400E

Nominal 4 foot (1.2m)

NU-543-500E

Nominal 5 foot (1.5m)

NU-543-600E

Nominal 6 foot (1.8m)

Performance Specifications

  1. Personal Protection

  2. Product Protection

NSF/ANSI 49

EN 12469

NSF/ANSI 49

EN 12469

NSF/ANSI 49

EN 12469

NSF/ANSI 49

EN 12469

NSF/ANSI 49

Class II

Class II

Class II

Class II


Style of Cabinet

Bench top/console w/basestand/storage

cabinet

Bench top/console w/base stand/storage

cabinet

Bench top/console w/base stand/storage

cabinet

Bench top/console w/base stand/storage

cabinet


Cabinet Construction

All welded stainless steel 16/18 gauge, Type 304

pressure tight design

All welded stainless steel 16/18 gauge, Type 304

pressure tight design

All welded stainless steel 16/18 gauge, Type 304

pressure tight design

All welded stainless steel 16/18 gauge, Type 304

pressure tight design

Diffuser for Air Supply (Metal)

Non-flammable

Non-flammable

Non-flammable

Non-flammable

HEPA Filter Seal Type:

Supply Filter-99.995% Eff. on MPPS

Exhaust Filter-99.995% Eff. on MPPS


HEPEX Seal

Neoprene, Spring-loaded


HEPEX Seal

Neoprene, Spring-loaded


HEPEX Seal

Neoprene, Spring-loaded


HEPEX Seal

Neoprene, Spring-loaded

Fumigation : per NIH/NSF Procedures

Yes

Yes

Yes

Yes

Standard Services:

Service Coupling (3/8 inch NPT) Service Coupling (3/8 inch NPT)

Gas Valve/Service Coupling (3/8inch NPT)

Duplex Outlet


One Right Sidewall Two Left Sidewall One Right Sidewall

Two Backwall


One Right Sidewall Two Left Sidewall One Right Sidewall

Two Backwall


One Right Sidewall Two Left Sidewall One Right Sidewall

Two Backwall


One Right Sidewall Two Left Sidewall One Right Sidewall

Two Backwall

Optional Services: Gas Cocks 3/8" NPT Ultraviolet Light Standard/Cup Sinks


Up to 3 ea. Sidewall One, Backwall

Left or Right Work

Surface


Up to 3 ea. Sidewall One, Backwall

Left or Right Work

Surface


Up to 3 ea. Sidewall One, Backwall

Left or Right Work

Surface


Up to 3 ea. Sidewall One, Backwall

Left or Right Work

Surface

Cabinet Size Inches (mm): Height (Fully Assembled)

Height (Minimum for Transport) Width

Depth with Armrest Removed


61 7/8 (1572)

59 (1499)

41 5/8 (1057)

31 7/16 (799)


61 7/8 (1572)

59 (1499)

53 5/8 (1362)

31 7/16 (799)


61 7/8 (1572)

59 (1499)

65 5/8 (1669)

31 7/16 (799)


61 7/8 (1572)

59 (1499)

77 5/8 (1972)

31 7/16 (799)

Work Access Opening Inches (mm): Standard Opening Height/Optional

Standard Inflow Velocity


10 (254) 105 FPM (.53

m/s)


10 (254)

105 FPM (.53 m/s)


10 (254)

105 FPM (.53 m/s)


10 (254)

105 FPM (.53 m/s)

Work Zone Inches (mm): Height

Width

Depth measured at 10 inches (254mm)

window height


28 1/2 (724)

34 3/8 (873)

25 3/4 (654)


28 1/2 (724)

46 3/8 (1178)

25 3/4 (654)


28 1/2 (724)

46 3/8 (1178)

25 3/4 (654)


28 1/2 (724)

70 3/8 (1788)

25 3/4 (654)

Viewing Window Inches (mm):

Standard is safety plate sliding glass

Fully closed to

21(533) open

Fully closed to

21(533) open

Fully closed to

21(533) open

Fully closed to

21(533) open

Required Exhaust CFM/CMH (opening) Standard/Optional:

Canopy Variable Flow Thimble (NU-911)

Canopy Fixed Flow Thimble (NU-907)

10(254)

CFM (CMH) 276-501 (649-851)

320 (544)

10(254)

CFM (CMH) 363-588 (617-1000)

426 (724)

10(254)

CFM (CMH) 451-676 (766-1149)

531 (902)

10(254)

CFM (CMH) 538-763 (915-1297)

634 (1077)

Plant Duct Static Pressure Eng./Metric

0.05-0.1"/1.27-2.54mm

H2O

0.05-0.1"/1.27-2.54mm

H2O

0.05-0.1"/1.27-2.54mm

H2O

0.05-0.1"/1.27-2.54mm

H2O

Heat Rejected, BTU, Per Hour (opening) (non-vented)

(vented)

10(254)

903

120

10(254)

1140

157

10(254)

1768

198

10(254)

1884

198

Electrical: (CE Marked) Volts, AC 50/60 Hz

+Amps: Blower/Lights (10 opening) Amps: Outlet

Rated Amps:

12 ft. Power Cord (one)


230

2.3

3

8

14 gauge - 3 Wire, 15A


230

2.9

3

10

14 gauge - 3 Wire, 15A


230

4.5

3

11

14 gauge - 3 Wire, 15A


230

4.8

3

11

14 gauge - 3 Wire, 15A

Crated Shipping Weight:***

Net Weight

460 lbs. /209 kg.

410 lbs. /186 kg.

530 lbs. /240 kg.

480 lbs. /218 kg.

620 lbs. /281 kg.

570 lbs. /258 kg.

690 lbs. /313 kg.

640 lbs. /290 kg.

Sound Pressure Level per ISO 4871****

Not to Exceed 55 dbA

Not to Exceed 56 dbA

Not to Exceed 58 dbA

Not to Exceed 60 dbA


***Crated shipping weight does not include weight for accessories or options

+ Based on cabinet with new filters running at 230VAC.

****Uncertainty is K = 2 dbA, measurement performed per ISO 11201 in normal running mode.

■Reference the customer test report for procedure and results.

Videos