[Instructions] Copper Water Pan NU-5860 Operation and Installation Instructions

NuAire Laboratory Equipment Instructions

Issue link: https://www.nuaire.com/learn/i/927153

Contents of this Issue

Navigation

Page 0 of 1

IN0039 | Rev 4 January/2018 Page 1 of 2 NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com ACCESSORY INFORMATION NU-5860 Copper Water Pan Operation and Installation Instructions For Series 5700, 5800 & 8600 Incubators NuAire recommends the use of this humidity pan for when copper components are desired as a part of the chamber contamination control. Place the pan on the bottom of the incubator chamber. The pan should be surface disinfected before usage with 70% alcohol or similar disinfectant. Fill with single deionized or distilled water (No purer than 1 mega ohm). No other additives are required for effective antimicrobial properties. See GTB0256 for care and use of copper. Note: Surface disinfectants that contain Halogens (i.e. Fluorine, chlorine, bromine, iodine, astatine) along with citric acid are not recommended as they cause discoloration. Copper Water Pan Location