More Views

LabGard® ES NU-430 Class II, Type B2 Biosafety Cabinet

The LabGard® ES (Energy Saver) NU-430 Total Exhaust Biological Safety Cabinet (BSC) may be used for work with non-flammable, non-explosive volatile chemicals and radionuclides by providing personnel, product, and environmental protection by exhausting 100% of contaminated air.


NSF/ANSI 49   UL Canada United States Listing

Part #: NU-430

Features

The LabGard® ES (Energy Saver) Class II, Type B2 Laminar Flow Biological Safety Cabinet is available as a bench or table top cabinet designed to provide optimum control over product quality in the lab, while also reducing the risk of exposing the product or lab personnel to any airborne biological or particulate chemical agents. This cabinet -- which meets all requirements of a Class II, Type B2 cabinet -- is equipped with an energy-saving motor to maximize energy efficiency and filter loading capacity. As prescribed by the Center for Disease Control in Atlanta, Georgia, the cabinet is intended for use in low to moderate risk-hazard research and drug preparation or product operations.


LabGard model NU-430 has been discontinued and replaced by LabGard model NU-560.

Options

Access Openings

Part Number Description
M243 8" (203 mm) access opening tempered sliding glass

Supply HEPA Filter

Part Number Description
M323 Wood Frame Supply HEPA Filter 99.99% efficient @ 0.3 Micron
M330 Metal Frame Supply H14 HEPA Filter 99.995% @ MPPS

Exhaust HEPA Filter

Part Number Description
M332 Wood Frame Exhaust HEPA Filter 99.99% efficient @ 0.3 Micron
M333 Metal Frame Exhaust H14 HEPA Filter 99.995% @ 0.3 MPPS

Filter Options

Part Number Description
M293 Supply Airflow Pre-filter
M295 Bag-In/Bag-Out Exhaust HEPA Filter
M294 Charcoal Exhaust Filter Module

Armrest

Part Number Description
M324 Heavy Duty Ergonomic PVC Armrest
M211 Stainless Steel Armrest

Plumbing Options

Part Number Description
M120 3/8" Soft Copper Pipe
M121 3/8" Black Pipe
M122 1/2" Rigid Copper Pipe

Decorative Panels

Part Number Description
M268 Flush Mount Front Panel
M237 Side Panel w/ Plumbing Out Bottom (Qty 2)
M238 Side Panel w/ Plumbing Out Top (Qty 2)
M239 Side Panel w/ Plumbing Out Back (Qty 2)
M284 Side Panel without Plumbing (Qty 2)

Built-in Options

Part Number Description
E34 Ultraviolet Light (115V)
M262 IV Bar w/ 6 Hooks
M428 NSF Cord Pass-Thru Left Side Wall
M427 NSF Cord Pass-Thru Right Side Wall
M387 Prop-up Work Tray
M411 Stainless Steel Window Edge Protector

Airflow Options

Part Number Description
E49 Air Supply On/Off Control (100V/115V)
E48 Cabinet Off Override
E51 Delay Off Exhaust Airflow
E50 Delay On Supply Airflow (100V/115V)
E47 Exhaust Airflow Loss Override

Outlets

Part Number Description
E59 15 Amp Type B Outlet (3 amp max) (Qty 1) (Left Front Farring) (115V)
E02 GFI Type B Outlet (3 amp max) (Qty 1) (Left Front Faring) (115V)
E05 15 Amp (3 Amp Max) Type B Outlet Front Faring (115V)
E64 GFI Type B Outlet (3 amp max) (Qty 1) (Left Front Faring), 15 Amp Type B Outlet (3 amp max) (Qty 1) (Right Front Faring)
E96 Side Wall Outlet (Qty 2) (Left/Right) (100V/115V)

Seismic Options

Part Number Description
M264 Seismic Brackets
M306 Seismic Brackets for Adjustable Base Stand

Service Valves

Part Number Description
M193 Left Top Service Valve
M194 Left Middle Service Valve
M195 Left Bottom Service Valve
M196 Right Top Service Valve
M197 Right Middle Service Valve
M198 Right Bottom Service Valve

Remote Controlled Service Valves

Part Number Description
M199 Left Top Remote Control Service Valve
M200 Left Middle Remote Control Service Valve
M201 Left Bottom Remote Control Service Valve
M202 Right Top Remote Control Service Valve
M203 Right Middle Remote Control Service Valve
M204 Right Bottom Remote Control Service Valve

Service Couplings

Part Number Description
M205 Left Top Service Coupling
M206 Left Middle Service Coupling
M207 Left Bottom Service Coupling
M208 Right Top Service Coupling
M209 Right Middle Service Coupling
M285 Right Bottom Service Coupling

Specs


Catalog Number

Catalog Number

NU-430-300

Nominal 3 foot (0.9m)

NU-430-400

Nominal 4 foot (1.2m)

NU-430-600

Nominal 6 foot (1.8m)

Performance Specifications

  1. Personal Protection

  2. Product Protection


NSF/ANSI 49


NSF/ANSI 49


NSF/ANSI 49

NSF Std. No. 49 Class

Class 11, Type B2

Class 11, Type B2

Class 11, Type B2

Style of Cabinet

Bench Top/Console w/Basestand/

Storage Cabinet

Bench Top/Console w/Basestand/

Storage Cabinet

Bench Top/Console w/Basestand/

Storage Cabinet

Cabinet Construction

All Welded Stainless Steel 16GA, Type 304 Pressure

ight Design

All Welded Stainless Steel 16GA, Type 304 Pressure

Tight Design

All Welded Stainless Steel 16GA, Type 304 Pressure

Tight Design

Diffuser for Air Supply (Metal)

Non-Flammable

Non-Flammable

Non-Flammable

HEPA Filter Seal Type:

Supply Filter-99.99% Eff. on 0.3 Microns Exhaust Filter-99.99% Eff. on 0.3 Microns


HEPEX Seal

Neoprene, Spring loaded


HEPEX Seal

Neoprene, Spring loaded


HEPEX Seal

Neoprene, Spring loaded

Fumigation per NIH/NSF Procedure

Yes

Yes

Yes

Standard Services:

Service Coupling (3/8 inch NPT)

Gas Valve/Service Coupling (3/8inch NPT) Duplex Outlet

One, Right Sidewall One, Right Sidewall One, Left Front Faring

One, Right Sidewall One, Right Sidewall One, Left Front Faring

One, Right Sidewall One, Right Sidewall One, Left Front Faring

Optional Services: Gas Cocks 3/8" NPT

Remote Controlled Valves** Ultraviolet Light

Standard/Cup Sinks


Up to 3 ea. Sidewall Up to 3 ea. Sidewall One, Backwall

Left or Right Work Surface


Up to 3 ea. Sidewall Up to 3 ea. Sidewall One, Backwall

Left or Right Work Surface


Up to 3 ea. Sidewall Up to 3 ea. Sidewall One, Backwall

Left or Right Work Surface

Cabinet Size Inches (mm): Height (Fully Assembled)

Height (Minimum for Transport) Width

Depth (with Control Center)


61 (1549)

61 (1549)

37 5/8 (956)

32 7/8 (835)


61 (1549)

61 (1549)

53 5/8 (1362)

32 7/8 (835)


61 (1549)

61 (1549)

77 5/8 (1972)

32 7/8 (835)

Work Access Opening Inches (mm): Standard Opening Height

Standard Inflow Velocity


8 (203)

105 FPM (.53 m/s)


8 (203)

105 FPM (.53 m/s)


8 (203)

105 FPM (.53 m/s)

Work Zone Inches (mm): Height

Width

Depth


25 1/2 (648)

30 3/8 (772)

23 1/2 (597)


25 1/2 (648)

46 3/8 (1178)

23 1/2 (597)


25 1/2 (648)

70 3/8 (1788)

23 1/2 (597)

Viewing Window Inches (mm): Standard is Tempered Sliding Glass

Hinged Tempered Glass (optional)

1.0” (25mm) Closed

18 1/2 (470) Open

8 (203) Access Opening

1.0” (25mm) Closed

18 1/2 (470) Open

8 (203) Access Opening

1.0” (25mm) Closed

18 1/2 (470) Open

8 (203) Access Opening

Certification Exhaust Value CFM/CMH

Concurrent Balance Value CFM/CMH+

496 ( 843)

546 (928)

754 (1281)

785 (1334)

1100 (1867)

1250 (2124)

Exhaust Duct Collar Dia. Inches (mm):

8 (203)

12 (305)

12 (305)

Plant Duct Static Pressure Eng/Metric

1.5” w.g. / 38mm w.g.

1.7” w.g. / 43mm w.g.

1.8” w.g. / 46mm w.g.

Heat Rejected,BTU,Per Hour

364

474

584

Electrical: 115V/(230V) Volts, AC (Hz)

++Amps: Blower/Lights Amps: Duplex

Amps: Total

12 ft. Power Cord (one)

U.L./U.L.-C (115V) 115, 60

1.8

3

8

14 GA - 3 Wire, 15A

U.L./U.L.-C (115V) 115, 60

2.1

3

8

14 GA - 3 Wire, 15A

U.L./U.L.-C (115V) 115, 60

2.6

3

10

14 GA-3 Wire, 15A Std.

Crated Shipping Weight:

Net Weight

510 lbs. / 209 kg.

460 lbs. / 231 kg.

570 lbs. / 259 kg.

520 lbs. / 236 kg.

760 lbs. / 345 kg.

710 lbs. / 322 kg.

**Remote controlled valve handles project through front faring. Decorative side panels are available to cover plumbing.

+Concurrent Balance Value shall be used for design and balance exhaust/supply HVAC requirements.

++Based on cabinet with new filters running at 115 VAC.