The PharmaGard NU-NTE800 Compounding Aspetic Containment Isolator (CACI) is a Total Exhaust Restricted Access Barrier System (RABS) designed to meet USP 800 requirements when compounding hazardous drugs. Model NU-NTE800 creates a negative-pressure ISO Class 5 clean-air environment to prevent contamination and protect personnel when compounding hazardous sterile preparations.
The large work zone lets workers place the interchange on the left or right sides of the 4-foot units, or on the left, right, or both sides at once of the 6-foot units. This provides each worker with a customized work space for increased comfort and reduced arm, neck and shoulder strain.
The oval slanted glove ports and the six-hook, three-position IV bar improve your reach while ensuring proper compounding distance and height.
Provide maximum knee and thigh clearance and proper height for optimal leg and upper-body support with the adjustable base.
The cool white lighting and open sight lines help workers see the work area more clearly.
Provide a certified ISO Class 5 sterile environment with this compounding aseptic containment isolator (CACI), which meets or exceeds the requirements of USP 797 for moving product in and out and for containing particulates and volatile gases. The HEPA-filtered interchange compartment maintains a negative pressure relative to the work area, which operates at a negative air pressure relative to the surrounding room. This minimizes the possibility of hazardous materials migrating from the workspace into the surrounding room, and of contamination migrating into the work zone.
Use this CACI, which meets USP 800 standards, to adequately protect personnel and the environment outside the pharmacy when handling hazardous drugs.
Protect workers from sharps and waste, which are disposed of through covered negative-pressure drop tubes. This waste system maintains sterility and increases workspace. A locking mechanism with an airtight seal holds the waste containers in place under the work surface, and the containers can be quickly released, removed and replaced.
The counterbalanced transfer window separates the inter-chamber from the sterile work zone. A security lock ensures the inner transfer door remains locked if the outer door is open. Workers can quickly move the transfer window up or down with one finger.
A removable stainless steel tray located in the transfer compartment is mounted on ball-bearing brackets to let workers smoothly move compounding materials into and out of the work area. The tray has coved corners to make it easier to keep clean and sterile.
Reduce energy costs and increase filter life with the ultrahigh-efficiency motor.
Catalog Number |
Catalog Number |
|
NU-NTE800-400 Nominal 4 foot (1.2m) |
NU-NTE800-600 Nominal 6 foot (1.8m) |
|
Performance Specifications 1. Product Protection |
CETA CAG-002-2006 ISO Class 5 (Laminar Flow) |
CETA CAG-002-2006 ISO Class 5 (Laminar Flow) |
Style of Isolator |
Bench top/console Isolator |
Bench top/console Isolator |
Isolator Construction |
Welded stainless steel 16GA, Type 304 pressure tight design |
Welded stainless steel 16GA, Type 304 pressure tight design |
Interchange Chamber |
ISO Class 5 With Internal/External Sealed Doors |
ISO Class 5 With Internal/External Sealed Doors |
Diffuser for Air Supply (Metal) |
Non-flammable |
Non-flammable |
HEPA Filter Seal Type: Supply Filter-99.99% Eff. on 0.3 microns Exhaust Filter-99.99% Eff. on 0.3 microns |
Neoprene, Spring loaded Under Negative Pressure |
Neoprene, Spring loaded Under Negative Pressure |
Standard Services: Hospital Grade Duplex Outlet |
One, Backwall |
One, Backwall |
Optional Services: Service Valves 3/8" NPT |
Up to 3 |
Up to 3 |
Isolator Size Inches (mm): Height Height (Center of Glove Port at 46-1/8) Width Depth (with Control Center) |
61 (1550) 92 1/2 (2350) 50 (1270) 32 1/2 (826) |
61 (1550) 92 1/2 (2350) 74 (1880) 32 1/2 (826) |
Work Zone Inches (mm): Width Height Depth (Center of Glove Port) |
35 1/4 (895) 27 3/8 (695) 23 1/4 (591) |
59 1/4 (1505) 27 3/8 (695) 23 1/4 (591) |
Interchange Inches (mm) Width Depth (at Work Surface) Height |
14 1/8 (359) 24 (610) 27 3/8 (695) |
14 1/8 (359) 24 (610) 27 3/8 (695) |
Hinged Lockable Viewing Window (Polycarbonate) |
Fully closed to fully open |
Fully closed to fully open |
Exhaust CFM/CMH Standard: Plan Duct Static Pressure Eng/Metric: (water gauge) |
550/934 2.0”w.g./50.8 mm w.g. |
725/1232 2.0”w.g./50.8 mm w.g. |
Heat Rejected, BTU, Per Hour |
539 |
880 |
Electrical: Volts, AC 60 Hz Amps: Blower/Lights Amps: Hospital Grade Duplex Amps: Total 12 ft. Hospital Grade Power Cord (one) |
UL/UL-C Listed 115 3 3 6 14 GA - 3 Wire, 15A |
UL/UL-C Listed 115 5 3 8 14 GA - 3 Wire, 15A |
*Crated Shipping Weight: *Net Weight: Net Optional Adjustable Automatic Base Stand Wt: Net Optional Telescoping Base Stand Wt: Net Optional Bench Mount Wt: |
515 lbs./234 kg. 465 lbs./211 kg. 150 lbs./68 kg. 60 lbs./27 kg. 20 lbs./19 kg. |
665 lbs./302 kg. 615 lbs./279 kg. 160 lbs./73 kg. 70 lbs./32 kg. 20 lbs./19 kg. |
* For total weight, must select and add (1) of (3) base options.