More Views

LabGard® ES NU-813 Bench Top Class I Containment Ventilated Enclosure (CVE)

The LabGard® ES (Energy Saver) model NU-813 is a Containment Primary Engineering Control (C-PEC) classified as a Class I Containment Ventilated Enclosure (CVE) offering personnel protection for the safe use of non-sterile hazardous drugs (HDs) in a Containment Segregated Compounding Area (C-SCA). The CVE may also be used for the use of animal necropsy.


UL Canada United States Listing

Part #: NU-813

Features

The LabGard® ES NU-813 Bench Top Class I Containment Ventilated Enclosure (CVE) is a benchtop containment cabinet intended to provide personnel protection when handling low-risk aerosols and particulate matter. To protect people working in the lab, the cabinet provides an air barrier, which minimizes the chance that any contaminants generated within migrate into the room. Equipped with a HEPA filter, the air flows through the cabinet then exits through the filter, which removes any contaminants that could put personnel at risk. This containment ventilated enclosure (CVE) is equipped with an electronic control system for easy usage, as well as an ultra-high-efficiency motor, which can help you save energy and cut energy bill costs in the lab.


Safety

Dynamic Air Barrier

The LabGard® ES Nu-813 Bench Top Class I Containment Cabinet is equipped with an air barrier for personnel protection. The cabinet has a strong inflow air velocity - at 80 fpm (.40 m/s), which minimizes the risk of contaminant migration.


Redundant Exhaust Filter Option

Redundant HEPA Filtration SystemThis containment cabinet has multiple filter options, which were created to fit the needs of your specific lab. You can either choose to use a Dual HEPA filter, which is recirculated into the lab environment, or you can opt for a single filter connected to your HVAC system. This option meets compounding pharmacy standard USP 800 for Non-sterile Hazardous powders.


Technology

Ultra High-Efficiency Motor

The cabinet is outfitted with an ultra-high-efficiency motor, which offers several benefits to users. The motor will help you save on your energy bills, increase filter life and ultimately lower your total cost of ownership.


Bag in Bag Out Exhaust HEPA filter(s)

Bag in bag out HEPA filter(s)The NU-813 cabinet was created to be convenient and easy for lab technicians to use. Its bag in bag out exhaust HEPA filter is a simplified way for you to remove the exhaust HEPA filter, without the risk of exposing yourself or other lab personnel to potential contaminants after decontamination procedures.


AeroMax Electronic Control System

AeroMax Electronic Control SystemThe AeroMax Control System makes the cabinet user-friendly and straightforward to control. The control system, which has an easy user interface, LED screen, and function keys, can perform a range of useful functions, including controlling the blower, controlling lights, controlling outlets, disabling the audible alarm switch, controlling blower DC ECM motor and more. The AeroMax Control System also monitors and displays airflow system performance via its PresurFlow monitor.


Parts & Accessories
Options

Filter Configuration

Part Number Description
M422 Exhaust HEPA Filter 99.99% @ 0.3 μ Efficient
M423 Redundant Exhaust HEPA Filter System 99.99% @ 0.3 μ Efficient (Qty 2)

Lighting Options

Part Number Description
E120 Fluorescent Lighting (115V)
E122 LED Lighting

Built-in Options

Part Number Description
M438 Lay In Charcoal Filter
M01 Service Valve (Qty 1)
M181 Seismic Studs
M449 Stainless Steel Work Surface Insert
M397 Waste Chute

Outlet Options

Part Number Description
E05 Outlet 115V (Please state location)
E02 GFI Type B Outlet (115V) (Right sidewall unless specified)

Specs


Catalog Number

Catalog Number

NU-813-300/E

Nominal 3 foot (0.9m)

NU-813-400/E

Nominal 4 foot (1.2m)

NU-813-600/E

Nominal 6 foot (1.5m)

Performance Specifications

  1. Personal Protection

  2. Product Protection

ANSI/ASHRAE 110

NSF/ANSI 49

Surrogate Powder

ANSI/ASHRAE 110

NSF/ANSI 49

Surrogate Powder

ANSI/ASHRAE 110

NSF/ANSI 49

Surrogate Powder

NSF Std. No. 49 Class

Class I

Class I

Class I

Style of Cabinet

Bench top/console with

optional base stand

Bench top/console with

optional base stand

Bench top/console with

optional base stand

Cabinet Construction

Polycarbonate panels with welded blower module.

Polycarbonate panels with welded blower module.

Polycarbonate panels with welded blower module.

HEPA Filter Seal Type: Exhaust Filter-99.995%

Eff. on 0.3 microns


Neoprene, Spring loaded


Neoprene, Spring loaded


Neoprene, Spring loaded

Fumigation per NIH/NSF

Procedure

Yes

Yes

Yes

Optional Services: Gas Cocks 3/8" NPT


Up to 3 ea. Sidewall


Up to 3 ea. Sidewall


Up to 3 ea. Sidewall

Cabinet Size Inches (mm): Height

Depth Width


54 (1372)

29 (737)

36 (914)


54 (1372)

29 (737)

48 (1219)


54 (1372)

29 (737)

72 (1829)

Work Access Opening Inches (mm):

Standard Opening Height/Optional

Standard Inflow Velocity


8 (203)

80 FPM (.40 m/s)


8 (203)

80 FPM (.40 m/s)


8 (203)

80 FPM (.40 m/s)

Work Zone Inches (mm): Height

Depth

Width


28 (711)

24 (610)

34-3/8 (873)


28 (711)

24 (610)

46-3/8 (1178)


28 (711)

24 (610)

70-3/8 (1788)


Viewing window is ¼ in (6mm). clear polycarbonate


Closed: 8 (203)

Open: 24 (610)


Closed: 8 (203)

Open: 24 (610)


Closed: 8 (203)

Open: 24 (610)

Noise: dbA Ambient

Running


53

57


53

57


53

62

CFM @ 80 FPM

195 CFM

260 CFM

390 CFM

Heat Rejected, BTU, Per Hour (non-vented)

265

330

600

Electrical: 115V (230V) Volts, AC (Hz)

Amps: Blower Amps: Lights Amps: Total

12 ft. Power Cord (one)


115 (230) / 50/60

1.6 (.8)

.2 (.2)

1.8 (1.0)

14 GA - 3 Wire, 15A


115 (230) / 50/60

1.8 (.8)

.2 (.2)

2.0 (1.0)

14 GA - 3 Wire, 15A


115 (230) / 50/60

3.2 (1.1)

.4 (.2)

3.6 (1.3)

14 GA - 3 Wire, 15A

Crated Shipping Weight: ***

Net Weight

350 lbs. / 159 kg.

300 lbs. / 136 kg.

450 lbs. / 204 kg.

400 lbs. / 181 kg.

550 lbs. / 249 kg.

500 lbs. / 227 kg.


***Crated shipping weight does not include weight for accessories or option

Download Resources
Reviews

Be the first to review this product

Write Your Own Review

Only registered users can write reviews. Please, log in or register