Model NU-560 features an exhaust interlock system that prevents operation of the internal supply blower unless the exhaust flow is sufficient to provide the correct air barrier inflow velocity at startup.
LabGard TE is a Total Exhaust Class II, Type B2 Biosafety Cabinet exhausting 100% of airflow through a facility’s HVAC system.
The TouchLink ECS is a color touch screen providing simple operation of cabinet functions.
A single ultra-high efficiency motor with optimally determine forward curved fan uses less energy to operate while containing more torque to extend the life of your HEPA filters.
Externally mounted, energy-efficient LED lighting provides 80 to 150 foot-candles (860 to 1600 LUX) on work surface.
At the core of any LabGard BSC is the cabinet’s shell. Robust sheets of stainless steel are bent, folded, and seam welded to create a pressure tight, a monolithic design minimizing the use of sealants.
A single piece of stainless steel is bent to create the interior work zone providing coved interior corners for easy cleaning.
5 Years Parts and Labor including Supply HEPA filter (United States and Canada)
6 Years Parts including Supply HEPA filter (International)
Access Opening |
|
560(Size)M443 | 8-Inch (203 mm) Access Opening Tempered Sliding Glass |
Supply Filter |
|
560(Size)M416 | Metal Frame Supply HEPA Filter 99.99% @ 0.3 Micron |
Exhaust Filter |
|
560(Size)M417 | Metal Frame Exhaust HEPA Filter 99.99% @ 0.3 Micron |
Bag-in Bag-Out Filter |
|
560(Size)M295 | Bag-in/Bag-out Exhaust HEPA Filter |
560(Size)M345 | No Bag-in/Bag-out Exhaust HEPA Filter |
Power Inputs |
|
560E07 | Junction Box |
560E60 | Power Cord (115V) (15 Amp) |
UV Light |
|
560(Size)E34 | Ultraviolet (UV) Light (115V) |
560(Size)E61 | No Ultraviolet (UV) Light (115V) |
Armrest |
|
560(Size)M211 | Stainless Steel Armrest |
560(Size)M324 | Heavy Duty Ergonomic Plastic Armrest |
Airflow Control |
|
560E47 | Exhaust Airflow Loss Override |
560E48 | Cabinet Off Override |
560E49 | Air Supply On/Off Control (100V/115V) |
560E50 | Delay On Supply Airflow (100V/115V) |
560E51 | Delay Off Exhaust Airflow |
Other |
|
430M427 | NSF Cord Pass-Thru Right Side Wall |
430M428 | NSF Cord Pass-Thru Left Side Wall |
560(Size)M262 | IV Bar with 6 Hooks, 3 Height Locations |
560(Size)M387 | Prop-Up Work Surface Tray |
560(Size)M411 | Stainless Steel Window Sash Edge Protector |
Outlets |
|
560E02 | GFI Type B Outlet (3 amp max) (Qty 1) (Left Front Faring) (115V) |
560E05 | 15 Amp Type B Outlet (3 amp max) (Qty 2) (Right & Left Front Farring) (115V) |
560E59 | 15 Amp Type B Outlet (3 amp max) (Qty 1) (Left Front Farring) (115V) |
560E68 | Outlet Cover Plate (Right Front Faring) |
Side Panels |
|
560M334 | No Side Panels |
560M284 | Side Panel without Plumbing (Qty 2) |
560M237 | Side Panel w/ Plumbing Out Bottom (Qty 2) |
560M238 | Side Panel w/ Plumbing Out Top (Qty 2) |
560M239 | Side Panel w/ Plumbing Out Back (Qty 2) |
Plumbing |
|
560M120 | 3/8" Soft Copper Pipe |
560M121 | 3/8" Black Pipe |
560M122 | 1/2" Rigid Copper Pipe |
Service Valves |
|
560M193 | Left Top Service Valve |
560M194 | Left Middle Service Valve |
560M195 | Left Bottom Service Valve |
560M196 | Right Top Service Valve |
560M197 | Right Middle Service Valve |
560M198 | Right Bottom Service Valve |
Remote Controlled Service Valves |
|
560M199 | Left Top Remote Control Service Valve |
560M200 | Left Middle Remote Control Service Valve |
560M201 | Left Bottom Remote Control Service Valve |
560M202 | Right Top Remote Control Service Valve |
560M203 | Right Middle Remote Control Service Valve |
560M204 | Right Bottom Remote Control Service Valve |
Service Couplings |
|
560M205 | Left Top Service Coupling |
560M206 | Left Middle Service Coupling |
560M207 | Left Bottom Service Coupling |
560M208 | Right Top Service Coupling |
560M209 | Right Middle Service Coupling |
560M285 | Right Bottom Service Coupling |
Seismic |
|
560(Size)M264 | Seismic Brackets |
560(Size)M306 | Seismic Brackets for Adjustable Bast Stand |
IQ/PQ/CQ |
|
NU-560-3Q | IQ/PQ/CQ Document Binder |
Catalog Number |
NU-560-400
|
NU-560-600
|
Performance Specifications |
||
1. Personal Protection 2. Product Protection |
NSF/ANSI 49 | NSF/ANSI 49 |
NSF Std. No. 49 Class | Class II, Type B2 | Class II, Type B2 |
Style of Cabinet | Bench Top/Console w/Base Stand/ Storage Cabinet |
Bench Top/Console w/Base Stand/ Storage Cabinet |
Cabinet Construction | All Welded Stainless Steel 16GA, Type 304 Pressure Tight Design |
All Welded Stainless Steel 16GA, Type 304 Pressure Tight Design |
Diffuser for Air Supply (Metal) | Non-Flammable | Non-Flammable |
HEPA Filter Seal Type: |
||
Supply Filter-99.99% Eff. on 0.3 Microns Exhaust Filter-99.99% Eff. on 0.3 Microns |
HEPEX Seal Neoprene, Spring loaded |
HEPEX Seal Neoprene, Spring loaded |
Fumigation per NIH/NSF Procedure | Yes | Yes |
Standard Services: |
||
Service Coupling (3/8 inch NPT) | Left Middle, Left Bottom, Right Middle | Left Middle, Left Bottom, Right Middle |
Gas Valve/Service Coupling (3/8 inch NPT) | Right Bottom | Right Bottom |
Duplex Outlet | One, Left Front Faring | One, Left Front Faring |
Optional Services: |
||
Gas Cocks 3/8" NPT | Up to 3 ea. Sidewall | Up to 3 ea. Sidewall |
Remote Controlled Valves* | Up to 3 ea. Sidewall | Up to 3 ea. Sidewall |
Ultraviolet Light | One, Backwall | One, Backwall |
Cabinet Size Inches (mm): |
||
Height (Fully Assembled) (Incl. Exh Filter Fastener) | 62 (1575) | 62 (1575) |
Height (Minimum for Transport) | 62 (1575) | 62 (1575) |
Width | 53 5/8 (1362) | 53 5/8 (1362) |
Depth (with Armrest removed) | 31 7/16 (798) | 31 7/16 (798) |
Work Access Opening Inches (mm): Standard Opening Height | 8 (203) | 8 (203) |
Standard Inflow Velocity | 105 FPM (0.53 m/s) | 105 FPM (0.53 m/s) |
Work Zone Inches (mm): |
||
Height | 25 3/16 (640) | 25 3/16 (640) |
Width | 46 3/8 (1178) | 70 3/8 (1788) |
Depth measured at 8 inches (203mm) | 26 (660) | 26 (660) |
Viewing Window Inches (mm): |
||
Standard is Tempered Sliding Glass Hinged Tempered Glass (optional) |
1.0 (25mm) Closed 18 1/2 (470) Open 8 (203) Access Opening |
1.0 (25mm) Closed 18 1/2 (470) Open 8 (203) Access Opening |
Certification Exhaust Value CFM/CMH | 740/1257 | 1122/1906 |
Concurrent Balance Value CFM/CMH + | 815/1385 | 1275/2166 |
Plant Duct Static Pressure Eng./Metric | 1.7 w.g./43mm w.g. | 1.8 w.g./46mm w.g. |
Heat Rejected, BTU, Per Hour | 474 | 584 |
Electrical: |
U.L./UL-C, US Listed 115 Volts, 60 Hz | U.L./UL-C, US Listed 115 Volts, 60 Hz |
Amps: Blower/Lights Amps: Outlet*** | 2.1 | 2.6 |
Rated Amps: | 8 | 10 |
12 ft. Power Cord (one) | 14 GA - 3 Wire, 15A | 14 GA - 3 Wire, 15A |
Crated Shipping Weight: | 540 lbs. / 245 kg. | 730 lbs. / 331 kg. |
Net Weight | 490 lbs. / 222 kg. | 680 lbs. / 308 kg. |
* Remote controlled valve handles project through front fairing. Decorative side panels are available to cover plumbing.
** Concurrent Balance Value shall be used for design and balance exhaust/supply HVAC requirements. Values provided are nominal. Design tolerances should be added for measurement and system differences.
*** Based on a cabinet with new filters running at 115 VAC.