Blizzard VFT NU-99579VFT -86°C Ultralow Freezer

Your samples require cold storage you can depend on around the clock. The Blizzard VFT NU-99579VFT delivers ENERGY STAR-certified -86°C ultralow temperature performance in a nominal 579-liter footprint, with Variable Frequency Technology that adapts to your lab's energy demands and 22-hour battery-backed alarm coverage when power is interrupted.
Blizzard VFT NU-99579VFT -86°C Ultralow Freezer, closed door front view
Your samples require cold storage you can depend on around the clock. The Blizzard VFT NU-99579VFT delivers ENERGY STAR-certified -86°C ultralow temperature performance in a nominal 579-liter footprint, with Variable Frequency Technology that adapts to your lab's energy demands and 22-hour battery-backed alarm coverage when power is interrupted.

Overview

When samples cannot be replaced, their storage environment must never fail. The Blizzard VFT NU-99579VFT is designed for laboratories that require reliable -86°C ultralow temperature performance in a compact, energy-efficient footprint. With a nominal 579-liter (20.4 cu. ft.) internal volume, it provides capacity for up to 40,000 two-milliliter vials, without the larger floor footprint of higher-capacity models.

Variable Frequency Technology (VFT) continuously modulates compressor speed to match real-time thermal demand. The result is ENERGY STAR-certified performance at 8.0 kWh per day. Two user-selectable operating modes adapt to your lab's current priority: High Performance mode for tightest temperature uniformity during active research periods, or Energy Saving mode for reduced power consumption during stable, long-term archiving.

Four independent inner door sections isolate each sample access event to a single compartment, protecting the remaining three zones from thermal disturbance. A 22-hour battery backup sustains alarm coverage, monitoring temperature, power status, door position, probe integrity, condenser condition, and battery state, throughout extended power interruptions. Performance data is available at any time via touchscreen LCD display, exportable by USB, and transmittable via RS-485 for integration with your laboratory monitoring system.

R290/R170 hydrocarbon refrigerant delivers near-zero global warming potential, meeting current environmental compliance requirements. The NU-99579VFT accepts 100–230V / 50–60Hz electrical input, compatible with laboratory environments worldwide. To request a quote or discuss your application, contact NuAire.

Frequently Asked Questions

What temperature does the Blizzard VFT NU-99579VFT reach?

The NU-99579VFT maintains temperatures from -40°C to -86°C with ±3°C temperature uniformity. Variable Frequency Technology continuously adjusts compressor output to sustain stable ultralow temperature storage.

What is the nominal capacity of the NU-99579VFT?

The NU-99579VFT has a nominal internal volume of 579 liters (20.4 cu. ft. nominal), storing up to 400 two-inch (5 cm) boxes across 16 racks, or approximately 40,000 two-milliliter vials.

Is the Blizzard VFT NU-99579VFT ENERGY STAR certified?

Yes. The NU-99579VFT is ENERGY STAR certified, with measured energy consumption of 8.0 kWh per day. Two selectable operating modes (High Performance and Energy Saving) allow labs to balance temperature precision and energy efficiency.

How long does battery backup last on the NU-99579VFT?

The NU-99579VFT provides 22 hours of battery-backed alarm coverage, maintaining active monitoring of temperature, power status, door position, probe integrity, and condenser condition during power interruptions.

What electrical supply does the NU-99579VFT require?

The NU-99579VFT accepts 100–230V / 50–60Hz input, compatible with laboratory electrical systems in North America, Europe, and worldwide without modification.

ENERGY STAR Certified: 8.0 kWh/day

Variable Frequency Technology adjusts compressor output in real time, delivering ENERGY STAR-certified energy performance at 8.0 kWh per day. Select High Performance mode for tightest temperature control or Energy Saving mode when efficiency is the priority.

Four Independent Inner Doors

Four thermally isolated inner door compartments limit temperature disturbance to the section being accessed. The remaining three compartments maintain stable conditions, protecting every sample that isn't being touched.

22-Hour Battery Backup

Battery-backed alarm coverage remains active for 22 hours through power events, monitoring temperature, power failure, door-ajar, probe failure, condenser condition, and battery state continuously.

Hydrocarbon Refrigerant: Near-Zero GWP

R290/R170 hydrocarbon refrigerant carries near-zero global warming potential, meeting current environmental compliance requirements and positioning labs for anticipated future refrigerant standards.

Universal Electrical Compatibility

100–230V / 50–60Hz input accepts a wide range of laboratory electrical environments without modification, from North America to Europe and international installations.

Specifications

Blizzard VFT NU-99579VFT -86°C Ultralow Freezer
2-inch Box Capacity 400
Rack Capacity 16
2 mL Vial Capacity 40,000
Rack Configuration 5x5
Shelves 3
Shelf Dimensions W x D Inches (mm) 23.9 x 25.6 (606 x 650)
Shelf Max. Capacity lbs (kg) 158.7 (72)
Cabinet Type Upright
Cabinet Thickness Inches (mm) 3.54 (90)
Porthole(s) 2
Outer Door Single Solid Door
Outer Door Thickness Inches (mm) 3.54 (90)
Outer Door Sealing Structure 2 outer sealing strips, 2 layers each (4 layers total)
Padlock Adapter Kit 1
Built-in Key Door Lock 1
Keys 4
Inner Doors 4 Solid Doors
Inside Door Sealing Structure 1 inner sealing strip, 1 layer
Shelves / Inner Doors 3 / 4
Adjustable Shelves Yes
Foot Yes
Casters Yes, 4
Stop Moving Feet Yes
Exterior Material Color-Coated Cold Plate Steel
Interior Material Color-Coated Steel Plate
Insulation VIP + CFC-Free Foaming Agent
Shelf Material Stainless Steel
Display Touch Screen LCD
Controller Level Top
Controller Type Microprocessor
Temperature Setting Mode Digital (Min. Accuracy 1°C)
Screen Setting Protection Yes (Password)
Setpoint Security Yes
Alarms High/Low Temperature, Adjustable High/Low Temperature, Ambient Temperature Display, High Ambient Temperature, Start-Up Delay, Sensor Error, Power Failure, High/Low Voltage, Voltage Exceed, Filter Dirty, Hot Condenser, Door Ajar, Low Battery
Data Output Features USB, Remote Alarm Terminals, RS-485
Power Switch 15A, On-Off with Circuit Breaker
Backup Battery Yes
Battery Runtime 22 Hours
Temp. Recorder Battery Runtime 6 Hours
Boot Delay Protection Yes
5V Power Supply Port Yes
Exterior Width Inches (mm) 35.2 (895)
Exterior Depth Inches (mm) 39.3 (998)
Exterior Height Inches (mm) 78.0 (1980)
Cabinet Width w/o Handle & Hinge Inches (mm) 31.5 (800)
Door Width w/o Handle Inches (mm) 32.5 (825)
Depth w/o Door, w/ Drawers Inches (mm) 33.9 (860)
Interior Width Inches (mm) 24.4 (620)
Interior Depth Inches (mm) 28.2 (716)
Interior Height Inches (mm) 51.6 (1310)
Usable Compartment Dimensions Inches (mm) 299 / 305 / 305 / 302
Packing Dimension W x D x H Inches (mm) 37.6 x 42.2 x 83.8 (953 x 1071 x 2127)
Container Load 20'/40'/40'H 12 / 24 / 24
Power Supply 100-230V, 50/60Hz
Power (W) 1100
Max Power Draw (W) 1100
Electrical Current (A) 14
Power Lead Length 2m
Power Consumption (Set -70°C @ 25°C ambient) 6.5 kWh/24h
Power Consumption (Set -80°C @ 25°C ambient) 7.5 kWh/24h
Power Consumption (Set -70°C @ 20°C ambient) 5.6 kWh/24h
Power Consumption (Set -80°C @ 20°C ambient) 8.3 kWh/24h
Power Consumption KWh/day 8.0
Sound Level 43.5 dB(A)
Temperature Range -40 to -86°C
Cooling Performance -86°C
Internal Temperature Uniformity ±3°C
Internal Temperature Accuracy 0.1°C
Sound Level 43.5 dBA
Safety Certification CE, UL, ENERGY STAR
Cooling Speed, 20°C Ambient (20°C down to -80°C) 239 Minutes
Holdover Time, 20°C Ambient (-80°C up to -50°C) 245 Minutes
1-Minute Door-Open Recovery to -75°C, 20°C Ambient 10 Minutes
1-Minute Door-Open Recovery to -75°C, 25°C Ambient 40 Minutes
Heat Rejection at 20°C Ambient 28,320 Btu/day
Heat Rejection at 25°C Ambient 25,591 Btu/day
Climate Class N
Cooling Type Direct Cooling
Compressor(s) 2
Size of Compressor 1 hp
Compressor Capacity 938W
Refrigerant HC
Refrigerant Model R290/R170
Temperature Sensor Thermistor Sensor
Controller Sensor PT100
Defrost Mode Manual
Net Weight lbs (kg) 716.5 (325)
Gross Weight lbs (kg) 782.6 (355)

Options

  • LUT-UFDR-552: Freezer Rack, Side Access, 5x5 with Qty 25 2-inch boxes
  • HAI-DCJ-55-B: Freezer Rack, Side Access, 5x5 for 2-inch boxes, Capacity 25
  • BIO-CO2BS-ARD: CO2 Back-Up Kit - Field Install, ship via truck only
  • BIO-LN2BS-ADDON: LN2 Back-Up Kit - Field Install, ship via truck only
  • HAI-UB2I: CO2 Back-Up Kit - Field Install, ship via ocean only
  • HAI-UBN2: LN2 Back-Up Kit - Field Install, ship via ocean only
  • HAI-960: Chart Recorder, 7 day, 6-inch Circular
  • HAI-219: Chart Recorder Replacement Paper, 52 Charts

Request Your Quote